A client is scheduled for an MRI of the lumbar spine without contrast. Which finding on the safety screening form requires the nurse to contact the MRI department before the client is transported?
- AThe client has an implanted cardiac pacemaker
- BThe client had a titanium hip prosthesis placed five years ago
- CThe client has several amalgam dental fillings
- DThe client's eGFR is 95 mL/min/1.73 m²
Show answer and rationale
Answer: A
A pacemaker in a strong magnetic field can be heated, displaced, reprogrammed or inhibited. Some newer devices are MRI-conditional, but only the MRI team can confirm the exact model, lead system and the conditions under which scanning is permitted, usually with cardiology present. The screening finding must therefore be communicated before the client leaves the unit.
Why the others are wrong: Modern orthopaedic prostheses are made of non-ferromagnetic alloys and are routinely scanned; they are noted on the form but do not stop the scan. Dental amalgam is not ferromagnetic. Kidney function matters for gadolinium contrast, and this scan is without contrast; an eGFR of 95 is normal in any case.
Source: Canadian Association of Radiologists MRI safety guidelines